EMS-1000 Viscometer
Power Point Presentation
Introducing a new approach to measuring viscosity!
Advantages
- Small Sample Volume
- Disposable Sample Container
- Sealed Vial for Measurement
From the esteemed Japanese manufacturer KEM (Kyoto Electronics Manufacturing Company, LTD) comes a new instrument that is at the forefront of viscosity testing technology.
The EMS-1000 Vicsometer provides contact-free measurement. The sample is placed in a disposable tube, and an aluminum sphere - also disposable - is dropped in. Inside the instrument, magnets surrounding the tube create a rotating magnetic field that drives the motion of the sphere within the sample.
The disposable tube and sphere allow for a sterile testing environment. There's no clean-up required at the end of the test. All you have to do is throw away the sample tube, preventing the possibility of cross-contamination between products or samples.
Measurement accuracy for toxic, volatile, or air-sensetive samples is ensured with the EMS-1000 Viscometer's sealed measuring cell. A CMOS camera also makes it easy to observe coagulation or detect impurities by letting you see the sample tube during the test.
Temperature control between 0 to 200oC is a built-in feature of the instrument. Tests can also be conducted at elevated pressures.
Product Specifications
Method of Measurement |
Electro Magnetically Spinning Method (EMS) |
Viscosity Measuring Range |
with 2mm sphere: 0.1-100,000 cP with 4.7mm sphere: 10-1,000,000 cP |
Measurement Accuracy and Repeatability | RSD 3% (according to KEM's standard measuring conditions) |
Minimum Sample Amount |
with 2mm sphere: 300µL with 4.77mm sphere: 700µL |
Temperature Range and Stability |
Range: 0-200oC (the ambient temperature must not exceed 20oC for measurements at temperatures less than 10o Stability: +/- 0.1oC |
Motor Speed Range and Stability |
Range: 500-1,000 rpm Stability: +/- 3% |
Ambient Conditions |
Temperature: 10-30oC Humidity: 85% RH max (no condensation) |
Storage Conditions |
Temperature: -20 - 60oC Humidity: 85% RH max (no condensation) |
Power Supply, main unit | AC 200-220V +/- 10% or AC 100-120V +/- 10% |
Power Consumption, main unit | Approx 150W (MAX 600 W) |
Dimensions | 374(W) x 347(D) x 335(H) mm |
Weight | Approx. 24kg |